Testomat 808 Heyl
2.999,00 €
Tilgængelig, leveringstid: 1-3 Tage
Produktinformation "Testomat 808 Heyl"
The Testomat® 808 from HEYL is a compact measuring device for automated online monitoring of water hardness. The measurement is based on the principle of limit value control with colour change and covers a range from 0.02 to 5 °dH (0.4 - 89 ppm).
Typical areas of application for permanent residual hardness control
- Reverse osmosis Systems
- Feed water treatment for boiler systems
- Provision of softened water for commercial applications
- Ultrapure water treatment
Water connections and operating conditions
- Operating pressure (depending on device version)
- 0.3 to 1 bar (0.3 x 10⁵ to 1 x 10⁵ Pa)
- 1 to 4 bar (1 x 10⁵ to 4 x 10⁵ Pa)
- For pressure ranges from 4 to 8 bar, an optionally available pressure reducer must be used.
- Water inlet: Opaque hose, 6 mm outer diameter / 4 mm inner diameter.
- Water outlet: Hose with 6 mm outer diameter / 4 mm inner Diameter.)
- •Water temperature: 10 to 40 °C
Functional features and equipment
- Optical display for limit value control using green or red LED
- Modern pump system for indicator supply
- Automated measuring process with low water consumption
- Clear display of error states and indicator quantities
- Manual control for internal and external flushing of the measuring chamber
- Option for controlling an external flushing Valve
- Three potential-free relays for:
- Controlling an external flushing valve
- Evaluating the measured value with regard to the set limit value
- External control or alarm systems
- Safety system for unattended operation for up to 72 Hours
- Use of 100 ml and 500 ml indicator bottles
- 4–20 mA interface for transmissible status and error messages
- RS-232 interface for firmware updates
Important note: For operating pressures above 4 bar, a suitable pressure reducer (sold separately) must be used. Please only use indicators approved for the Testomat® 808 or Testomat® BOB.
Power supplies: 230 VAC (180-270 VAC)
Flow rate: 3,5 l/h @ 1,5 bar
Max. Suction height: 1.5 m
Ambient temperature: 0 - 45 °C
Temperature dosing medium: 0 - 50 °C
Installation class: II
Stroke frequency / strokes per minute: 0 - 180
Noise class: 74 dBA
Temperature transport and storage: -10 to 50°C (14 to 122 °F)
IP - protection class: IP 65
Switching capacity optional "alarm" relay: max. 2 A / 230 VAC, 0.5A / 24 VDC
Gebrüder Heyl Vertriebsgesellschaft für innovative Wasseraufbereitung mbH
Max-Planck-Str.
31135
Hildesheim
Deutschland
vertrieb@heylneomeris.de